LARMOR and ZOOM seem generally very happy with IBEX's performance. Below is the feedback from the sprint demo meeting with them, loosely categorised:
- Rob would like an update on progress!
- As the scaling on live view changes between (for example) 4.9 and 4.91 it changes the width of the plot which gives the impression of a "judder".
- Rob would like some instructions on how to set up a 1D live view.
- Time-of-flight cut-off is not working properly - the data does not look like Rob expects. Apparently mantid has the same bug.
- LARMOR would like to display two live views simultaneously, each of which is pinned to a specific period number
- XY profile is "ridiculously useful" for aligning a beam-stop
- Wiring table readback in the GUI is not updating when the tables are changed via genie_python
- Alarms and ioc log perspectives "never contain anything of use" so they don't look at them
- Python console can pop up in the wrong view
- Would like script, graph and synoptic all on one screen
- How do we define the order of blocks when those blocks are defined in components
- LARMOR are still using
globals.txt
for most things. Asks what the "recommended" approach is. - If we set things up in the GUI configuration screens, can we see a list of ALL the macros that have been set (simultaneously)
- Would like some form of warning/protection against loading a configuration with inconsistent macros (e.g. clashing COM port settings). Can we run configchecker locally on a config change and warn if tests fail?
- LARMOR would like to refactor configurations and would like some advice about how best to set these up.
- LARMOR would like a way to "dismiss" a block even though it is defined in the current config, to save space. This might not be needed after the refactoring mentioned above.
- Long block names are being truncated in blocks view in the GUI
- The server is too slow to start ("multiple minutes")
- Lack of microsoft patches on NDH machine for processes not dying contributed to a DAE problem on surf/made it harder to diagnose.
- SURF still runs SECI but has some IBEX processes installed. As a user it is not clear exactly what I need to start/restart after an NDX power cycle
- Julabo - would like a way to send fixed values on IOC startup
- Would like a (better) way to set motor limits as part of a configuration
- KEPCO driver did not work. Failed to communicate. When the physical unit was swapped, the IOC talked but couldn't set a negative value