Original file line number Diff line number Diff line change
Expand Up @@ -394,14 +394,17 @@ To change this, we used the command "5 cca x = 34" and power cycled the controll
Now we can see that the piezo is set to the correct range (Pf).
With that verified, now confirm that the voltage output from the PCI Board is working:
1. Put the BNC cable input currently in EXT IN on the Tiger control panel into the input of the oscilloscope instead.
2. Go to the test panels for the PCI board in NI MAX.

| 1. Put the BNC cable input currently in EXT IN on the Tiger control panel into the input of the oscilloscope instead.
| 2. Go to the test panels for the PCI board in NI MAX.
.. image:: Images/TestPanels.png
:align: center
:alt: How to find the Test Panels panel
3. Set the voltage mode to sinewave generation.
4. Set the voltage range to be between 0 to 10 V.
5. Set the frequency to a desired value (we ended up setting it pretty high at 10000 Hz for ease of viewing on the oscilloscope).

| 3. Set the voltage mode to sinewave generation.
| 4. Set the voltage range to be between 0 to 10 V.
| 5. Set the frequency to a desired value (we ended up setting it pretty high at 10000 Hz for ease of viewing on the oscilloscope).
.. image:: Images/TestPanelConfiguration.png
:align: center
:alt: How to find the Test Panels panel
Expand Down