fix(api): default to both sensors on lld #16247
Merged
Add this suggestion to a batch that can be applied as a single commit.
This suggestion is invalid because no changes were made to the code.
Suggestions cannot be applied while the pull request is closed.
Suggestions cannot be applied while viewing a subset of changes.
Only one suggestion per line can be applied in a batch.
Add this suggestion to a batch that can be applied as a single commit.
Applying suggestions on deleted lines is not supported.
You must change the existing code in this line in order to create a valid suggestion.
Outdated suggestions cannot be applied.
This suggestion has been applied or marked resolved.
Suggestions cannot be applied from pending reviews.
Suggestions cannot be applied on multi-line comments.
Suggestions cannot be applied while the pull request is queued to merge.
Suggestion cannot be applied right now. Please check back later.
Overview
In hardware testing we were explicitly calling liquid_probe with a defined probe, but the protocol engine wasn't doing this and it was always defaulting to just the A1 nozzle. The expected behavior is that both sensors should work in an OR configuration. This just adds a check that if the # of nozzles > 1 to turn both on instead of just the A1 sensor.
Test Plan and Hands on Testing
Changelog
Review requests
Risk assessment