docs(api): updated Flex default flow rates #13796

Merged
merged 1 commit into from
Oct 18, 2023
Merged
Changes from all commits
Commits
File filter

Filter by extension

Filter by extension

Conversations
Failed to load comments.
Loading
Jump to
Jump to file
Failed to load files.
Loading
Diff view
Diff view
46 changes: 28 additions & 18 deletions api/docs/v2/new_pipette.rst
Original file line number Diff line number Diff line change
Expand Up @@ -345,11 +345,11 @@ If you know that all your liquid handling will take place in a specific mode, th
Pipette Flow Rates
==================

Measured in µL/s, the flow rate determines how much liquid a pipette can aspirate, dispense, and blow-out. Opentrons pipettes have their own default flow rates. The API lets you change the flow rate on a loaded :py:class:`.InstrumentContext` by altering the :py:obj:`.InstrumentContext.flow_rate` properties listed below.
Measured in µL/s, the flow rate determines how much liquid a pipette can aspirate, dispense, and blow out. Opentrons pipettes have their own default flow rates. The API lets you change the flow rate on a loaded :py:class:`.InstrumentContext` by altering the :py:obj:`.InstrumentContext.flow_rate` properties listed below.
Copy link
Contributor

Choose a reason for hiding this comment

The reason will be displayed to describe this comment to others. Learn more.

"blow out" oops, my bad. 🥇


* Aspirate: ``InstrumentContext.flow_rate.aspirate``
* Dispense: ``InstrumentContext.flow_rate.dispense``
* Blow-out: ``InstrumentContext.flow_rate.blow_out``
* Blow out: ``InstrumentContext.flow_rate.blow_out``

These flow rate properties operate independently. This means you can specify different flow rates for each property within the same protocol. For example, let's load a simple protocol and set different flow rates for the attached pipette.

Expand Down Expand Up @@ -394,17 +394,27 @@ These flow rates will remain in effect until you change the ``flow_rate`` attrib
Flex Pipette Flow Rates
-----------------------

The following table provides data on the default aspirate, dispense, and blow-out flow rates (in µL/s) for Flex pipettes.
The default flow rates for Flex pipettes depend on the maximum volume of the pipette and the capacity of the currently attached tip. For each pipette–tip configuration, the default flow rate is the same for aspirate, dispense, and blowout actions.

.. list-table::
:header-rows: 1

* - Pipette Model
- Tip Capacity (µL)
- Flow Rate (µL/s)
* - 50 µL (1- and 8-channel)
- All capacities
- 57
* - 1000 µL (1-, 8-, and 96-channel)
- 50
- 478
* - 1000 µL (1-, 8-, and 96-channel)
- 200
- 716
* - 1000 µL (1-, 8-, and 96-channel)
- 1000
- 716

+-------------------------------+-----------------+-----------------+-----------------+
| Pipette Models | Aspirate (µL/s) | Dispense (µL/s) | Blow-out (µL/s) |
+===============================+=================+=================+=================+
| | **Flex 50 µL pipettes** | 8 | 8 | 4 |
| | 1- and 8-channel | | | |
+-------------------------------+-----------------+-----------------+-----------------+
| | **Flex 1000 µL pipettes** | 160 | 160 | 80 |
| | 1-, 8-, and 96-channel | | | |
+-------------------------------+-----------------+-----------------+-----------------+

Additionally, all Flex pipettes have a well bottom clearance of 1 mm for aspirate and dispense actions.

Expand All @@ -413,7 +423,7 @@ Additionally, all Flex pipettes have a well bottom clearance of 1 mm for aspirat
OT-2 Pipette Flow Rates
-----------------------

The following table provides data on the default aspirate, dispense, and blow-out flow rates (in µL/s) for OT-2 GEN2 pipettes. Because the flow rates are the same across all three actions, we've consolidated the data into the "Flow Rates" column.
The following table provides data on the default aspirate, dispense, and blowout flow rates (in µL/s) for OT-2 GEN2 pipettes. Default flow rates are the same across all three actions.

.. list-table::
:header-rows: 1
Expand All @@ -422,25 +432,25 @@ The following table provides data on the default aspirate, dispense, and blow-ou
- Volume (µL)
- Flow Rates (µL/s)
* - P20 Single-Channel GEN2
- 1-20
- 120
-
* API v2.6 or higher: 7.56
* API v2.5 or lower: 3.78
* - P300 Single-Channel GEN2
- 20-300
- 20300
-
* API v2.6 or higher: 92.86
* API v2.5 or lower: 46.43
* - P1000 Single-Channel GEN2
- 100-1000
- 1001000
-
* API v2.6 or higher: 274.7
* API v2.5 or lower: 137.35
* - P20 Multi-Channel GEN2
- 1-20
- 120
- 7.6
* - P300 Multi-Channel GEN2
- 20-300
- 20300
- 94

Additionally, all OT-2 GEN2 pipettes have a default head speed of 400 mm/s and a well bottom clearance of 1 mm for aspirate and dispense actions.