66 changes: 55 additions & 11 deletions api/docs/v2/pipettes/partial_tip_pickup.rst
Original file line number Diff line number Diff line change
Expand Up @@ -29,8 +29,8 @@ Before getting started with partial tip pickup, make sure your protocol specifie
- Single, partial column
- 2.20

Nozzle Layout
=============
Nozzle Layouts
==============

Use the :py:meth:`.configure_nozzle_layout` method to choose how many tips a pipette will pick up. The method's ``style`` parameter accepts special layout constants. You must import these constants at the top of your protocol, or you won't be able to configure the pipette for partial tip pickup.

Expand All @@ -46,8 +46,8 @@ For greater convenience, also import the individual layout constants that you pl

Then when you call ``configure_nozzle_layout`` later in your protocol, you can use the shorter ``style=COLUMN``.

Column Layout Example
---------------------
Column Layout
-------------

Here is the start of a protocol that imports the ``COLUMN`` and ``ALL`` layout constants, loads a 96-channel pipette, and sets it to pick up a single column of tips.

Expand Down Expand Up @@ -97,8 +97,8 @@ In this configuration, pipetting actions will use a single column::

:py:meth:`.InstrumentContext.pick_up_tip` always accepts a ``location`` argument, regardless of nozzle configuration. Do not pass a value that would lead the pipette to line up over more unused tips than specified by the current layout. For example, setting ``COLUMN`` layout and then calling ``pipette.pick_up_tip(tip_rack["A2"])`` on a full tip rack will lead to unexpected pipetting behavior and potential crashes.

Row Layout Example
------------------
Row Layout
----------

Here is the start of a protocol that imports the ``ROW`` and ``ALL`` layout constants, loads a 96-channel pipette, and sets it to pick up a single row of tips.

Expand Down Expand Up @@ -135,12 +135,12 @@ You can also set ``start="A1"`` to use the backmost nozzles and pick up from the

.. note::

Consider the placement of your tip rack when choosing the ``start`` value for row pickup. The pipette cannot pick up from back to front (``start="H1"``) on tip racks in row A of the deck, nor can it pick up from front to back (``start="A1"``) on tip racks in row D of the deck. This is because the pipette would have to move too far backward or forward, respectively, to align over those tips.
Consider the placement of your tip rack when choosing the ``start`` value for row pickup. The pipette cannot pick up from back to front (``start="H1"``) on tip racks in row A of the deck, nor can it pick up from front to back (``start="A1"``) on tip racks in row D of the deck. This is because the pipette can't move far enough backward or forward, respectively, to align over those tips.

Use a different ``start`` value, or load the tip rack in row B or C.

Single Layout Example
---------------------
Single Layout
-------------

Single-tip pickup is available on both 8-channel and 96-channel pipettes. For 8-channel pipettes, there are two possible configurations, using either the front or back nozzle. For 96-channel pipettes, there are four possible configurations, using any of the corner nozzles.

Expand Down Expand Up @@ -221,8 +221,8 @@ Since this configuration uses ``start="H12"``, it will pick up tips in the usual
You can pick up tips row-wise first, rather than column-wise first, by specifying a location for :py:meth:`.pick_up_tip` each time you use it in ``SINGLE`` configuration. However, as with all partial tip layouts, be careful that you don't place the pipette in a position where it overlaps more tips than intended.


Partial Column Layout Example
-----------------------------
Partial Column Layout
---------------------

Partial column pickup is available on both 8-channel and 96-channel pipettes. Partial columns contain 2 to 7 consecutive tips in a single column. 8-channel pipettes always pick up partial columns with the frontmost nozzles of the pipette (``start="H1"``). 96-channel pipettes can pick up partial columns with the frontmost nozzles in the leftmost column (``start="H1"``) or the rightmost column (``start="H12"``).

Expand Down Expand Up @@ -250,6 +250,50 @@ To specify the number of tips to pick up, add the ``end`` parameter when calling
- C
- B

Here is the start of a protocol that imports the ``PARTIAL_COLUMN`` and ``ALL`` layout constants, loads an 8-channel pipette, and sets it to pick up four tips::

from opentrons import protocol_api
from opentrons.protocol_api import PARTIAL_COLUMN, ALL

requirements = {"robotType": "Flex", "apiLevel": "2.20"}

def run(protocol: protocol_api.ProtocolContext):
partial_rack = protocol.load_labware(
load_name="opentrons_flex_96_tiprack_1000ul",
location="B2"
)
trash = protocol.load_trash_bin("A3")
pipette = protocol.load_instrument("flex_8channel_1000", mount="left")
pipette.configure_nozzle_layout(
style=PARTIAL_COLUMN,
start="H1",
end="E1",
tip_racks=[partial_rack]
)

.. versionadded:: 2.20

This configuration will pick up tips from the back half of column 1, then the front half of column 1, then the back half of column 2, and so on::

pipette.pick_up_tip() # picks up A1-D1 from tip rack
pipette.drop_tip()
pipette.pick_up_tip() # picks up E1-H1 from tip rack
pipette.drop_tip()
pipette.pick_up_tip() # picks up A2-D2 from tip rack

When handling liquids in partial column configuration, remember that *the frontmost channel of the pipette is its primary channel*. For example, to use the same configuration as above to transfer liquid from wells A1–D1 to wells A2–D2 on a plate, you must use the wells in row D as the source and destination targets::

pipette.transfer(
volume=100,
source=plate["D1"], # aspirate from A1-D1
dest=plate["D2"], # dispense into A2-D2
)

.. warning::

Do not move the pipette to row A of labware when in ``PARTIAL_COLUMN`` configuration! This is different than when pipetting to a full column, either with the 8-channel pipette in ``ALL`` configuration or with the 96-channel pipette in ``COLUMN`` configuration.

If you pipette to row A, the frontmost tip will move to row A and the other tips — further back in the layout — will hang over the back edge of the labware. They will not enter the correct wells and they will likely cause a crash.