Analog settling & integration timing #7

Open
patricktokeeffe opened this issue Sep 18, 2020 · 0 comments
Open

Analog settling & integration timing #7

patricktokeeffe opened this issue Sep 18, 2020 · 0 comments
Labels
help wanted Extra attention is needed question Further information is requested

Comments

@patricktokeeffe
Copy link
Member

patricktokeeffe commented Sep 18, 2020

Non-default values are being used in several locations for Settling Time and Analog Integration Time. It's unclear if these represent arbitrary choices, or are optimal selections based on lab experiment, or are hold-overs from converting between logger types.

Group A (CR1000X)

  • fine wire thermocouples
    • settling time: 450 vs 500 (def)
    • fN1: 250 vs 15000 or _60Hz
  • panel temp
    • fN1: 250 vs 15000 or _60Hz
  • component (4 channel) radiometer, incl. self-temp
    • fN1: 15000 (= fast integration, 0.0667 millisecond) ✓
  • PAR sensor
    • fN1: 15000 ✓
  • T/RH probes (x4), via mux
    • fN1: 15000 ✓

Group B (CR1000X)

  • fine wire thermocouples
    • settling time: 450 vs 500 (def)
    • fN1: 250 vs 15000 or _60Hz
  • panel temp
    • fN1: 250 vs 15000 or _60Hz
  • PAR
    • fN1: 60 (=60 Hz noise rejection) ✓
  • T/RH probes (x4), via mux
    • fN1: 60 ✓

Group C (CR1000X)

  • fine wire thermocouples
    • settling time: 450 vs 500 (def)
    • fN1: 250 vs 15000 or _60Hz
  • panel temp
    • fN1: 250 vs 15000 or _60Hz
  • T/RH probes (x4), via mux
    • fN1: 15000 ✓ (declared as Long)

Group S (CR6)

  • panel temp
    • fN1: 60 ✓
  • soil averaging thermocouple
    • fN1: 60 ✓
  • soil heat flux plate, non-self-calibrating (x?)
    • fN1: 60 ✓
  • infrared surface thermometer, self & target temps
    • fN1: 60 ✓
  • soil heat flux plate, self-calibrating (x?) including heater
    • fN1: 60 ✓
  • PAR sensor
    • fN1: 60 ✓
@patricktokeeffe patricktokeeffe added help wanted Extra attention is needed question Further information is requested labels Sep 18, 2020
@patricktokeeffe patricktokeeffe pinned this issue Sep 18, 2020
Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment
Labels
help wanted Extra attention is needed question Further information is requested
Projects
None yet
Development

No branches or pull requests

1 participant