@zmgao did you update SDM delay steps for any of the LI-7500A/RS? I only had time to connect to the D2 IRGA last week and I recall seeing user-defined delay was still 0
The high-frequency instruments all have different inherent delays which must be accounted when aligning time-series data for flux computations.
CSAT3B User Manual rev 3/18
recommended delay step is 17 to yield total delay of 297ms
recommended delay step is 25 to yield total delay of 297ms
*need to verify, might be set to 17 already from prior use
**need to verify, not sure if these were ever changed