Skipped time series records #10

Open
patricktokeeffe opened this issue Sep 18, 2020 · 4 comments
Open

Skipped time series records #10

patricktokeeffe opened this issue Sep 18, 2020 · 4 comments
Labels
bug Something isn't working

Comments

@patricktokeeffe
Copy link
Member

Relatively infrequent, but there are skipped records in data table Time_Series indicating program timing problem. The GPS instruction inside the 10Hz scan should be moved to slow scan instead.

@patricktokeeffe patricktokeeffe changed the title Skipped time series records in Group B Skipped time series records Sep 18, 2020
@patricktokeeffe
Copy link
Member Author

Also observing this in Group C

@patricktokeeffe patricktokeeffe added the bug Something isn't working label Sep 18, 2020
@zmgao
Copy link
Contributor

zmgao commented Sep 19, 2020

I moved the GPS to slow scan, but there are still skipped records in data table Time_series.

@patricktokeeffe
Copy link
Member Author

SDM timing (the one LI7500 not using group trigger) might be the reason there are still skipped records.

Is there a commit or PR to reference for the changes?

@zmgao
Copy link
Contributor

zmgao commented Sep 23, 2020

There are skipped records for all three CR1000X.